NSN 6515-01-352-9507

Part Details | OXYGEN POWERED TIME CYCLED VOLUME VENTILATOR

6515-01-352-9507 An item designed to provide controlled ventilation for non-breathing patients and assisted ventilation for spontaneously breathing patients in hospital, ambulance, field uses and in military aeromedical aircraft. It may consist of a pneumatically controlled, manual and automatic selectable modes, portable ventilator apparatus time/volume cycled, airway pressure limited, patient valve, pressure relief valve, adjustable flow control valve, oxygen regulator with preset 50 psi outlet and DISS pin-indexed yoke to accommodate size D oxygen cylinder, hoses and other related features. Excludes VENTILATOR, POCKET, PORTABLE and VENTILATOR, VOLUME, PORTABLE.

Alternate Parts: 01FG5100, 6515-01-352-9507, 01-352-9507, 6515013529507, 013529507

Supply Group (FSG) NSN Assigned NIIN Item Name Code (INC)
65FEB 04, 199201-352-950749743 ( VENTILATOR, VOLUME, TIME CYCLED, OXYGEN POWERED )
REFERENCE DRAWINGS & PICTURES
Cross Reference | NSN 6515-01-352-9507
Part Number Cage Code Manufacturer
01FG510038186O-TWO MEDICAL TECHNOLOGIES INC
Technical Data | NSN 6515-01-352-9507
Characteristic Specifications
SPECIAL FEATURESPNEUMATICALLY CONTROLLED; MANUAL AND AUTOMATIC SELECTABLE MODES; IN CARRYING CASE; ALSO ACCOMODATES 1 SIZE D OXYGEN CYLINDER; MANUAL TRIGGER PRESET AT 40 LPM; AUTOMATIC MODE PROVIDES RATIO OF INSPIRATORY TO EXPIRATORY (I:E) OF 1:2 W/SELECTABLE BREATH PER MINUTE (BPM) FROM 24 BPM AT 0.3 LITER TIDAL VOLUME TO 12 BPM AT 1.2 LITERS TIDAL VOLUME WHICH AUTOMATICALLY ADJUST THE FLOW RATE; AIRWAY PRESSURE RELIEF VALVE PRESET AM; INCL PATIENT VALVE, PRESSURE RELIEF VALVE, ADJUSTABLE FLOW CONTROL VALVE, VENTILATOR, 50 PSI OUTLET AND DISS PIN-INDEXED YOKE FOR SIZE D 02 CYLINDER, AND HOSES; FOR EMERGENCY USE W/THE NONBREATHING AND SPONTANEOUS BREATHING PATIENT REQUIRING ASSISTED VENTILATION
PART NAME ASSIGNED BY CONTROLLING AGENCYVENTILATOR VOLUME TIME CYCLED OXYGEN POWERED EMERGENCY USE