NSN 6515-01-379-7523

Part Details | BLOOD OXYGENATOR

6515-01-379-7523 An item used for carbon dioxide removal from blood and replenishment of oxygen extracorporeally during cardiovascular surgery. It may consist of mixing chamber, reservoirs, heat exchangers, connectors, ports, filters, and the like. Excludes BLOOD FILTER UNIT, INDIRECT TRANSFUSION.

Alternate Parts: 0199701, 01997-01, ULTROX I, 6515-01-379-7523, 01-379-7523, 6515013797523, 013797523

Supply Group (FSG) NSN Assigned NIIN Item Name Code (INC)
65SEP 08, 199301-379-752338715 ( OXYGENATOR, BLOOD )
REFERENCE DRAWINGS & PICTURES
Cross Reference | NSN 6515-01-379-7523
Part Number Cage Code Manufacturer
01997-010X0G8AVECOR CARDUIBASCULAR
ULTROX I0X0G8AVECOR CARDUIBASCULAR
Technical Data | NSN 6515-01-379-7523
Characteristic Specifications
ACCESSORY COMPONENTS AND QUANTITYRECIRCULATION, PURGE/SAMPLE PORT, 1; ARTERIAL OUTLET, 1; ARTERIAL TEMPERATURE, MONITORING ADAPTOR, 1; GAS INLET, 1; MOUNTING BLOCKS, 2; OXYGENATOR, 1; GAS OUTLET, 1; WATER PORTS INLET/OUTLET, 1; HEAT EXCHANGER, 1; PRESSURE MONITORING LUER PORT, 1; VENOUS BLOOK INLET
FEATURES PROVIDED STERILE
SPECIAL FEATURESSURGICAL OXGENATOR WITH BUILT IN HEAT EXCHANGER; PURE SILICONE SOLID MEMBRANE; MEMBRANE SURFACE 3.5 M2; STATIC PRIMING VOLUME, 620 ML NOMINAL; TYPICAL PATIENT SIZING, UP TO 160 KGS; RECOMMENDED MAXIMUM BLOOD FLOW 6.5 LITERS/MIN; RECOMMENDED MAXIMUM GAS FLOW 15.0 LITERS/MIN; USED DURING ROUTINE CARDIOPULMONARY BYPASS PROCEDURES