Part Details | BLOOD OXYGENATOR
6515-01-382-4256 An item used for carbon dioxide removal from blood and replenishment of oxygen extracorporeally during cardiovascular surgery. It may consist of mixing chamber, reservoirs, heat exchangers, connectors, ports, filters, and the like. Excludes BLOOD FILTER UNIT, INDIRECT TRANSFUSION.
Alternate Parts: 16310, 16310, 6515-01-382-4256, 01-382-4256, 6515013824256, 013824256
| Supply Group (FSG) | NSN Assigned | NIIN | Item Name Code (INC) |
|---|---|---|---|
| 65 | NOV 02, 1993 | 01-382-4256 | 38715 ( OXYGENATOR, BLOOD ) |
REFERENCE DRAWINGS & PICTURES
Cross Reference | NSN 6515-01-382-4256
| Part Number | Cage Code | Manufacturer |
|---|---|---|
| 16310 | 76381 | 3M COMPANYDBA 3M |
| 16310 | 69892 | 3M COMPANYDIV HEALTH CARE BUSINESS |
Technical Data | NSN 6515-01-382-4256
| Characteristic | Specifications |
|---|---|
| COMPONENT AND QUANTITY | MICROPOROUS POLYPROPYLENE MEMBRANE; STAINLESS STEEL HEAT EXCHANGER; ACRYLIC CASE; POLYURETHANE AND SILICONE POTTING COMPOUNDS |
| SPECIAL FEATURES | BELOWS DESIGN; STERILE; NON-PYROGENIC; DISPOSABLE; SINGLE USE ONLY; PRIMING VOLUME 320 ML; HEAT EXCHANGE PERFORMANCE FACTOR, 0.58; FIBER SURFACE AREA, 1.82 M2; RECOMMENDED FLOW RANGE, 1-7 L/MIN; OXYGEN TRANSFER, VOLUME %, 5.7 ML/MIN: 341: CARBON DIOXIDE TRANSFER, VOLUMES %: 7.3 ML/MIN: 442 |