Part Details | BLOOD OXYGENATOR
6515-01-461-6370 An item used for carbon dioxide removal from blood and replenishment of oxygen extracorporeally during cardiovascular surgery. It may consist of mixing chamber, reservoirs, heat exchangers, connectors, ports, filters, and the like. Excludes BLOOD FILTER UNIT, INDIRECT TRANSFUSION.
Alternate Parts: MAXFTC, MAX-FTC, 6515-01-461-6370, 01-461-6370, 6515014616370, 014616370
| Supply Group (FSG) | NSN Assigned | NIIN | Item Name Code (INC) |
|---|---|---|---|
| 65 | DEC 22, 1998 | 01-461-6370 | 38715 ( OXYGENATOR, BLOOD ) |
REFERENCE DRAWINGS & PICTURES
Cross Reference | NSN 6515-01-461-6370
| Part Number | Cage Code | Manufacturer |
|---|---|---|
| MAX-FTC | 0ZDT9 | MEDTRONIC CARDIOPULMONARY |
Technical Data | NSN 6515-01-461-6370
| Characteristic | Specifications |
|---|---|
| ACCESSORY COMPONENTS AND QUANTITY | O2 IN PORT; PLASMA RESISTANT FIBERS; HIGH EFFICIENCY FIBER HEAT EXCHANGER; ARTERIAL SAMPLING; BOTTOM-OUT ARTERIAL BLOOD OUTLET; ARTERIAL TEMPERATURE PORT; RECIRCULATION PORT; WATER PORTS, 2; GENTLY-ANGLED VENOUS INLET |
| FEATURES PROVIDED | DISPOSABLE AND STERILE |
| SPECIAL FEATURES | MAXIMA FORTE HOLLOW FIBER OXYGENATOR WITH PLASMA RESISTANT FIBER AND INTEGRAL HARDSHELL VENOUS RESERVOIR; 1 TO 7 LITERS PER MINUTE BLOOD FLOW RATE; 295 ML STATIC PRIMING VOLUME; MICROPOROROUS POLYPROPYLENE HOLLOW FIBER MEMBRANE; 2.4 M2 MEMBRANE SURFACE AREA; PORT SIZES: 3/8 INCH VENOUS INLET, 3/8 INCH ARTERIAL OUTLET, 1/4 INCH RECIRCULATION, 1/4 INCH GAS IN-BARBED, 1/4 INCH GAS OUT-SMOOTH, FEMALE LUER ARTERIAL SAMPLE, 1/2 INCH QUICK DISCONNECT WATER IN AND OUT; 40 PSI MAXIMUM WATER PRESSURE |