Part Details | CARBON DIOXIDE GAS DETECTOR
6515-01-464-9877 A non-toxic, chemically treated indicator device, designed for use between a TUBE, TRACHEAL, and a breathing circuit. It is to be used for detecting the approximate range of expired respiratory carbon dioxide gas by a patient during tracheal intubation or while the patient is intubated while on mechanical ventilation. This item can be visible through a plastic covering, which allows it to change color with each breath without a delay during a patient's respiratory inspiration and expiration cycle.
Alternate Parts: 982467, 98-2467, PEDICAP, PEDI-CAP, 6515-01-464-9877, 01-464-9877, 6515014649877, 014649877
| Supply Group (FSG) | NSN Assigned | NIIN | Item Name Code (INC) |
|---|---|---|---|
| 65 | MAY 18, 1999 | 01-464-9877 | 47277 ( DETECTOR, CARBON DIOXIDE GAS ) |
REFERENCE DRAWINGS & PICTURES
Cross Reference | NSN 6515-01-464-9877
| Part Number | Cage Code | Manufacturer |
|---|---|---|
| 98-2467 | 89875 | DEFENSE LOGISTICS AGENCYDIRECTORATE OF MEDICAL MATERIEL |
| PEDI-CAP | 7Y988 | NELLCOR PURITAN BENNETT INCORPORATEDDBA COVIDIEN |
Technical Data | NSN 6515-01-464-9877
| Characteristic | Specifications |
|---|---|
| MATERIAL AND LOCATION | PLASTIC OVERALL |
| SPECIAL FEATURES | WT. 1 KG TO 15 KG; INTERNAL VOLUME: 3 CC; RESISTANCE TO FLOW: 2.5 CM H20 PLUS/MINUS 0.5 CM @ 10 L/MIN FLOW; DETECTOR WT.: LESS THAN 5 G; CONNECTOR PORTS, PATIENT END: 18 MM OD/15 MM ID; CIRCUIT END: 15 MM OD/5MM ID; USAGE TIME: UP TO 2 HOURS; RESPONDS QUICKLY TO EXHALED CO2 WITH A SIMPLE COLOR CHANGE FROM PURPLE TO YELLOW |
| SUPPLEMENTARY FEATURES | END-TIDAL; SIZE: PEDIATRIC |
| PART NAME ASSIGNED BY CONTROLLING AGENCY | DETECTOR CARBON DIOXIDE GAS 24S |