Part Details | SIMULATOR,OXIMETER
6515-01-476-9961
Alternate Parts: SP02INDEX2PFE, SP02INDEX2FE, 6515-01-476-9961, 01-476-9961, 6515014769961, 014769961
| Supply Group (FSG) | NSN Assigned | NIIN | Item Name Code (INC) |
|---|
REFERENCE DRAWINGS & PICTURES
Cross Reference | NSN 6515-01-476-9961
| Part Number | Cage Code | Manufacturer |
|---|---|---|
| SP02INDEX2PFE | 56198 | BIOTEK INSTRUMENTS, INC.DBA BIOTEK |
| SP02INDEX2FE | 56198 | BIOTEK INSTRUMENTS, INC.DBA BIOTEK |
Technical Data | NSN 6515-01-476-9961
| Characteristic | Specifications |
|---|---|
| GENERAL CHARACTERISTICS ITEM DESCRIPTION | CAN TEST THE OXIMETER W/O THE PROBE IN-LINE;PROBE ANALYSIS INCLUDES: LED TEST - VOLTAGE DROP ACROSS BOTH RED AND INFRARED DIODES, PHOTODIODE TEXT DETECTOR RESPONSE TO A 1MA ILLUMINATING CURRENT AND RESISTANCE BETWEEN ALL WIRES;CAN SIMULATE ARTERIAL OXYGEN SATURATION FROM 35-100 PCT OXYGENATION;SIMUULATES PATIENT CONDITIONS SUCH AS NORMAL, WEAK PULSE, GERIATRIC, OBESE, BRADYCARDIA AND TACHYCARDIA; PORTABLE;9 PRELOAD MANUFACTURERS R-CURVES;5 DOWNLOADALE R-CURVES;PHYSIOLOGICAL FINGER FOR SP02 TEST;RS232 PORT; PHYSIOLOGICAL MOTION ARTIFACT;HEARTRATE RANGE: 30-250 BPM: RESOLUTION: 1 BPM;PULSE AMPLITUDE RANGE: 0-20 PCT PERFUSION, RESOLUTION: 1 PCT; 2-LINE X 24-CHARACTER SUPER TWIST LCD DISPLAY |