NSN 6515-01-561-2008

Part Details | ELECTROCARDIOGRAPH AND CARDIAC MONITOR TESTER

6515-01-561-2008 A portable, battery-powered, device which simulates both normal and abnormal heart rhythms and when connected to an electrocardiograph or cardiac monitor, determination can be made as to whether these equipment items are producing accurate readings when used on a patient. A normal rhythm and various common arrhythmias provide for testing the recognition capabilities and proper operation of rate alarms and detection protocol.

Alternate Parts: 2558932, 6515-01-561-2008, 01-561-2008, 6515015612008, 015612008

Supply Group (FSG) NSN Assigned NIIN Item Name Code (INC)
65FEB 27, 200801-561-200836779 ( TESTER, ELECTROCARDIOGRAPH AND CARDIAC MONITOR )
REFERENCE DRAWINGS & PICTURES
Cross Reference | NSN 6515-01-561-2008
Part Number Cage Code Manufacturer
255893234511FLUKE BIOMEDICAL CORPORATION
Technical Data | NSN 6515-01-561-2008
Characteristic Specifications
SHAPE RECTANGULAR
SPECIAL FEATURESPOWER, 9 V ALKALINE BATTERY FOR 200-HOUR LIFE OR LINE-OPERATED VIA BATTERY ELIMINATOR; 115 VAC OR 230 VAC OR 7.7 VDC; 100 MA, UNREGULATED; CONNECTOR, 2.5 MM CENTER (+); DIMENSIONS, 5.2 IN L X 3.9 IN W X 1.4 IN H; WEIGHT, 0.9 LB; TEMPERATURE, OPERATING: 59 F TO 95 F DEGREES AND STORAGE 32 F TO 131 FDEGREES; 12 ARRHYTHMIA SELECTIONS: A TRIAL FIBRILLATION, SECOND-DEGREE A-V BLOCK, TYPE 1, RIGHT-BUNDLE-BRANCH BLOCK, PREMATURE ATRIAL CONTRACTION, PREMATURE VENTRICULAR CONTRACTION EARLY, PREMATURE VENTRICULAR CONTRACTION R ON T, MULTIFOCAL PVCS, BIGEMINY, RUN OF 5PVCS, VENTRICULAR TACHYCARDIA, VENTRICULAR FIBRILLATION, PACED; 9 V BATTERY
FUNCTIONAL DESCRIPTIONTHE PS400 IS A BASIC PATIENT SIMULATOR FOR QUALITY-ASSURANCE TESTING OFECG MACHINES, MONITORS AND TELEMETRY UNITS
END ITEM IDENTIFICATIONPS400 ECG/ARRHYTHMIA SIMULATOR