Part Details | TESTER,DEFIBRILLATO
6515-01-586-7368
Alternate Parts: 18249, 6625NCM104681, 2811919, IMPULSE 7000DP, 6515-01-586-7368, 01-586-7368, 6515015867368, 015867368
| Supply Group (FSG) | NSN Assigned | NIIN | Item Name Code (INC) |
|---|
REFERENCE DRAWINGS & PICTURES
Cross Reference | NSN 6515-01-586-7368
| Part Number | Cage Code | Manufacturer |
|---|---|---|
| 18249 | 89875 | DEFENSE LOGISTICS AGENCYDIRECTORATE OF MEDICAL MATERIEL |
| 6625NCM104681 | 64616 | DEFENSE MEDICAL STANDARDIZATIONBOARD |
| 2811919 | 34511 | FLUKE BIOMEDICAL CORPORATION |
| IMPULSE 7000DP | 34511 | FLUKE BIOMEDICAL CORPORATION |
Technical Data | NSN 6515-01-586-7368
| Characteristic | Specifications |
|---|---|
| OVERALL LENGTH | 13.000 INCHES NOMINAL |
| OVERALL WIDTH | 9.500 INCHES NOMINAL |
| OVERALL HEIGHT | 5.000 INCHES NOMINAL |
| SPECIAL FEATURES | LCD DISPLAY; INTERNAL RECHARGEABLE NIMH BATTERY PACK; MANUAL AND REMOTE MODES OF OPERATION; PACER BRAND SELECTIONS; TEN INDEPENDENT ECG OUTPUTS PROVIDE 12 LEAD COMBINATIONS FOR STANDARDIZED CLINICAL SIGNALS; FLEXIBLE HEART RATE SETTINGS; AED TECHNOLOGY COMPATIBILITY; ENERGY OUTPUT: PULSE TRIGGER LEVEL, 20 V; PULSE WIDTH RANGE: 1 MS TO 50 MS; VOLTAGE RANGE: 20 V TO 5000 V; CURRENT: 0.4 A TO 100 A; FREQUENCY RANGE: 2000 HZ TO 8000 HZ |
| STORAGE TEMP RANGE | -4.0 TO +140.0 DEG FAHRENHEIT |
| FUNCTIONAL DESCRIPTION | ENSURES PROPER OPERATION AND ULTIMATE PERFORMANCE OF CRITICAL LIFE-SUPPORT CARDIAC-RESUSCITATION EQUIPMENT |
| PART NAME ASSIGNED BY CONTROLLING AGENCY | TESTER DEFIBRILLATOR AND EXTERNAL PACER ANALYZER IMPULSE 7000DP |
| UNPACKAGED UNIT WEIGHT | 6.600 POUNDS |