Part Details | TRANSDUCER,BLOOD PR
6515-01-599-8568
Alternate Parts: 25273, 6515NCM116428, CPJ840J6, 6515-01-599-8568, 01-599-8568, 6515015998568, 015998568
| Supply Group (FSG) | NSN Assigned | NIIN | Item Name Code (INC) |
|---|
REFERENCE DRAWINGS & PICTURES
Cross Reference | NSN 6515-01-599-8568
| Part Number | Cage Code | Manufacturer |
|---|---|---|
| 25273 | 89875 | DEFENSE LOGISTICS AGENCYDIRECTORATE OF MEDICAL MATERIEL |
| 6515NCM116428 | 64616 | DEFENSE MEDICAL STANDARDIZATIONBOARD |
| CPJ840J6 | 0ZBJ4 | PHILIPS ELECTRONICS NORTH AMERICACORPORATION |
Technical Data | NSN 6515-01-599-8568
| Characteristic | Specifications |
|---|---|
| SPECIAL FEATURES | TRANSDUCER, BLOOD PRESSURE, INVASIVE; PIEZORESISTIVE; REUSABLE; INCLUDES: 3 METER (118 IN) LONG CABLE, DISPOSABLE SILICONE DOME, TRANSDUCER, TRANSDUCER HOLDER; FOR CONTINUOUS ARTERIAL OR VENOUS BLOOD PRESSURE MEASUREMENT; PRESSURE RANGE: -20 TO +300 MM HG; MAX OVERPRESSURE: 1000 MM HG; SENSITIVITY: 50 MICRO V/V/CM HG; RESONANCE FREQ: 300 HZ TYPICAL WITH TRANSDUCER DOME; MAX ELECTRICAL EXITATION: 15 V DC OR AC; INPUT RESISTANCE: 700 OHM DC AND 350 OHM AC; OUTPUT RESISTANCE: 1000 OHMS; OPERATING TEMPERATURE: 10 TO 50 DEG C; NON-LINEARITY AND HYSTERESIS: MAX 0.5 PCT OF FULL SCALE; ZERO BALANCE: MAX +/- 30 MM HG; THERMAL SENSITIVITY SHIFT: 0.15 PCT/DEG C; THERMAL ZERO SHIFT: MAX 0.15 MM HG/DEG C |
| STORAGE TEMP RANGE | -20.0 TO +70.0 DEG CELSIUS |
| END ITEM IDENTIFICATION | 6515-01-586-5646 MONITOR, PATIENT VITAL SIGNS |
| PART NAME ASSIGNED BY CONTROLLING AGENCY | TRANSDUCER AND CABLE BLOOD PRESSURE CPJ840J6 |