Part Details | LARYNGEAL AIRWAY MASK
6515-01-649-3369 An item used for insertion into the pharynx for better airway management and ease of patient monitoring. Comes in various sizes for all patients.
Alternate Parts: 329 150 000U, 1365643, 6515NCM152432, 6515-01-649-3369, 01-649-3369, 6515016493369, 016493369
| Supply Group (FSG) | NSN Assigned | NIIN | Item Name Code (INC) |
|---|---|---|---|
| 65 | OCT 27, 2015 | 01-649-3369 | 52503 ( MASK, AIRWAY, LARYNGEAL ) |
REFERENCE DRAWINGS & PICTURES
Cross Reference | NSN 6515-01-649-3369
| Part Number | Cage Code | Manufacturer |
|---|---|---|
| 329 150 000U | 65891 | AMBU INC. |
| 1365643 | 89875 | DEFENSE LOGISTICS AGENCYDIRECTORATE OF MEDICAL MATERIEL |
| 6515NCM152432 | 64616 | DEFENSE MEDICAL STANDARDIZATIONBOARD |
Technical Data | NSN 6515-01-649-3369
| Characteristic | Specifications |
|---|---|
| MATERIAL | PLASTIC |
| SIZE DESIGNATION | INFANT |
| USAGE DESIGN | ENDOTRACHEAL INTUBATION |
| FEATURES PROVIDED | DISPOSABLE AND STERILE |
| SPECIAL FEATURES | SIZE 1.5; PACKAGING COLOR CODE (GRAY); PATIENT WEIGHT 5-10 KG; MAXIMUM CUFF INFLATION VOLUME 7ML; MAXIMUM INTRA-CUFF PRESSURE 60 CM H2O; AIRWAY CONNECTOR 15MM ISO 5356-1; MIN. I.D. TUBE 6.9MM; INFLATION VALVE LURE CONE ISO 594-1; APPROPRIATE STORAGE TEMPERATURE +10 TO +25C (50-77F); WEIGHT 15G; INTERNAL VOLUME OF VENTILATOR PATHWAY 6ML; PRESSURE DROP <0.8CM H2O AT 15 L/MIN; MIN INTERDENTAL GAP 17MM; INTERNAL PATHWAY 10.7CM; MAX ET-TUBE 4.0; AIRWAY TUBE,CUFF, AIRWAY CONNECTOR, PILOT BALLOON, PILOT TUBE DEHP-FEE AND PVC; INFLATION VALVE DEHP FREE PVC/SILICONE |
| PART NAME ASSIGNED BY CONTROLLING AGENCY | MASK,AIRWAY,LARYNGEAL,DISPOSABLE, AURA-I, SIZE 1.5 |