Part Details | RADIOGRAPHIC AND FLUOROSCOPIC X-RAY APPARATUS
6525-00-965-2501
Alternate Parts: GGX635D, GG-X-635D, 6525-00-965-2501, 00-965-2501, 6525009652501, 009652501
| Supply Group (FSG) | NSN Assigned | NIIN | Item Name Code (INC) |
|---|---|---|---|
| 65 | JAN 01, 1960 | 00-965-2501 | 06321 ( X-RAY APPARATUS, RADIOGRAPHIC AND FLUOROSCOPIC ) |
REFERENCE DRAWINGS & PICTURES
Cross Reference | NSN 6525-00-965-2501
| Part Number | Cage Code | Manufacturer |
|---|---|---|
| GG-X-635D | 81348 | FEDERAL SPECIFICATIONSPROMULGATED BY GENERAL SERVICES |
Technical Data | NSN 6525-00-965-2501
| Characteristic | Specifications |
|---|---|
| DESIGN TYPE | LX ANY ACCEPTABLE |
| TUBE TYPE | DOUBLE FOCUS, ROTATING ANODE RADIOGRAPHIC AND DOUBLE FOCUS, ROTATING ANODE FLUOROSCOPIC |
| TUBE QUANTITY | A1$$A1 |
| TUBE COOLING METHOD | AIR AND AIR |
| TUBE MOUNTING TYPE | TUBE STAND AND RD ANY ACCEPTABLE |
| TUBE STAND MOUNTING TYPE | OVERHEAD |
| TABLE TYPE | TILTING |
| TILT RANGE | 90 DEG TRENDELENBURG TO VERTICAL RANGE OF POSITIONS |
| DRIVE TYPE | MOTOR |
| BUCKY DIAPHRAGM | INCLUDED |
| CONTROL TYPE | VERTICAL |
| MOUNTING TYPE | STATIONARY |
| OPERATION METHOD | AUTOMATIC |
| RADIOGRAPHIC TIMER TYPE | EXPOSURE |
| FLUOROSCOPIC TIMER TYPE | SYNCHRONOUS MOTOR |
| TRANSFORMER RECTIFICATION METHOD | ELECTRON TUBE |
| RECTIFICATION TYPE | FULL-WAVE |
| TRANSFORMER OUTPUT CAPACITY | 300 MA OUTPUT AT 150 KV PEAK |
| FLUOROSCOPIC SCREEN TYPE | ANY ACCEPTABLE |
| SCREEN LENGTH | 12.000 INCHES MINIMUM |
| SCREEN WIDTH | 12.000 INCHES MINIMUM |
| SCREEN TYPE | W/SPOT-FILM |
| CASSETTE SIZE | 8 IN. BY 10 IN.,10 IN. BY 12 IN. |
| VOLTAGE IN VOLTS | 220.0 |
| FREQUENCY IN HERTZ | 60.0 |
| PHASE | SINGLE |
| ELECTRIC POWER RATING | 40000.0 WATTS NOMINAL |
| ACCESSORY COMPONENTS AND QUANTITY | REMOTE CONTROLLED POWER DRIVEN BODY SECTIONAL DEVICE 1,COLLIMATOR 1,COMPRESSION DEVICE 1,FLUOROSCOPIC IMAGE INTENSIFIER W/22 CM DIA PHORPHOR 1,FOOTREST 1,ADJUSTABLE PATIENT GRASP HANDLE 2,TABLE TILT INDICATOR 1,PHOTO-TIMER RECEIVER IN SPOT FILM DEVICE 1,PHOTO-TIMER RECEIVER IN THE BUCKEY AND/OR CASSETTE CHANGER /OPTIONAL/ 1,RADIATION SHIELD 1,MYELOGRAPHIC PATIENT SUPPORT 1,SHOULDER SUPPORT 1 |
| SPECIAL FEATURES | FLOOR MOUNTED TABLE W/MOTOR DRIVEN ADJ TOP; DESIGNED TO USE W/MOTION PICTURE AND TELEVISION CAMERAS; REQUISITION SHOULD STATE WHETHER WALL-MOUNTED,HIGH VOLTAGE TAKE UP REEL IS REQUIRED |
| SUPPLEMENTARY FEATURES | REQUISITION SHOULD STATE FLOOR-LOADING CAPACITY,CEILING STRENGTH AND HEIGHT,WALL AND FLOOR CONSTRUCTION (CONCRETE,WOOD,ETC),ELECTRICAL SUPPLY AND CAPACITY,DIAGRAM OF ROOM INDICATING DIMENSIONS,COMPATABILITY W/EXISTING FACILITIES,AND WHETHER OR NOT CONTRACTOR INSTALLATION SERVICE IS REQUIRED |
| PART NAME ASSIGNED BY CONTROLLING AGENCY | X-RAY APP RAD&FLUOR300MA 150PKV 2 TUBE 230V 60HZ AC W/ROT ANODE |