Part Details | MEDICAL X-RAY APPARATUS TUBE UNIT
6525-01-012-9382 Consists of a tube insert in a suitable housing equipped with cable receptacles; without cables.
Alternate Parts: 652500C959115, S21919, 6525-01-012-9382, 01-012-9382, 6525010129382, 010129382
| Supply Group (FSG) | NSN Assigned | NIIN | Item Name Code (INC) |
|---|---|---|---|
| 65 | NOV 20, 1975 | 01-012-9382 | 15750 ( TUBE UNIT, MEDICAL X-RAY APPARATUS ) |
REFERENCE DRAWINGS & PICTURES
Cross Reference | NSN 6525-01-012-9382
| Part Number | Cage Code | Manufacturer |
|---|---|---|
| 652500C959115 | 81923 | ARMY MEDICAL SERVICEC/O THE SURGEON GENERAL DEPT OF THE |
| S21919 | 37480 | RAYTHEON COMACHLETT LABORATORIES INC MW AND PT |
Technical Data | NSN 6525-01-012-9382
| Characteristic | Specifications |
|---|---|
| PORT TO CABLE RECEPTACLE ANGLE IN DEG | 105.0 |
| TUBE INSERT ANODE TYPE | STANDARD SPEED ROTATING |
| TARGET BASE MATERIAL | TUNGSTEN |
| TUBE INSERT RADIOGRAPHIC PEAK VOLTAGE RATING | 125.0 KILOVOLTS |
| FOCAL SPOT QUANTITY | 2 |
| FOCAL SPOT SIZE | 1.0 MM SQ; 2.0 MM SQ |
| TARGET ANGLE IN DEG | 15.0 |
| FEATURES PROVIDED | EXPLOSION PROOF |
| ANODE HEAT CAPACITY RATING IN HEAT UNITS | 140000 |
| HOUSING X-RAY PROTECTIVE VALUE | X-RAY PROTECTIVE VALUE OF HOUSING SHALL BE IN ACCORDANCE WITH DHEW PUBLICATION NO. FDA 75-8003 JUL 1974 |
| CONTAINER TYPE | ANY ACCEPTABLE |
| PART NAME ASSIGNED BY CONTROLLING AGENCY | TUBE UNIT MEDICAL X-RAY APPARATUS TUNGSTEN 125 KILOVOLTS |