Part Details | MEDICAL X-RAY APPARATUS TUBE UNIT
6525-01-020-8352 Consists of a tube insert in a suitable housing equipped with cable receptacles; without cables.
Alternate Parts: 46155400G31, 46-155400G31, D1177G18, 6525-01-020-8352, 01-020-8352, 6525010208352, 010208352
Supply Group (FSG) | NSN Assigned | NIIN | Item Name Code (INC) |
---|---|---|---|
65 | MAY 07, 1976 | 01-020-8352 | 15750 ( TUBE UNIT, MEDICAL X-RAY APPARATUS ) |
REFERENCE DRAWINGS & PICTURES
Cross Reference | NSN 6525-01-020-8352
Part Number | Cage Code | Manufacturer |
---|---|---|
46-155400G31 | 24456 | GENERAL ELECTRIC COMEDICAL SYSTEMS DIV |
D1177G18 | 24456 | GENERAL ELECTRIC COMEDICAL SYSTEMS DIV |
Technical Data | NSN 6525-01-020-8352
Characteristic | Specifications |
---|---|
PORT TO CABLE RECEPTACLE ANGLE IN DEG | 180.0 |
TUBE INSERT ANODE TYPE | HIGH SPEED ROTATING |
TARGET BASE MATERIAL | MOLYBDENUM |
TARGET FACE MATERIAL | TUNGSTEN-RHENIUM ALLOY |
TUBE INSERT RADIOGRAPHIC PEAK VOLTAGE RATING | 150.0 KILOVOLTS |
FOCAL SPOT QUANTITY | 2 |
FOCAL SPOT SIZE | 1.0 MM SQ; 2.0 MM SQ |
TARGET ANGLE IN DEG | 15.0 |
ANODE HEAT CAPACITY RATING IN HEAT UNITS | 350000 |
HOUSING X-RAY PROTECTIVE VALUE | X-RAY PROTECTIVE VALUE OF HOUSING SHALL BE IN ACCORDANCE WITH DHEW PUBLICATION NO. FDA 75-8003 JUL 1974 |
CONTAINER TYPE | ANY ACCEPTABLE |
PART NAME ASSIGNED BY CONTROLLING AGENCY | TUBE UNIT MEDICAL X-RAY APPARATUS HI-SPEED ROTATING ANODE 150PKV |