Technical Data | NSN 6525-01-507-1623
| Characteristic | Specifications |
|---|---|
| SPECIAL FEATURES | SUITABLE FOR FIXED AND FIELD MILITARY MTF; 280VAC + OR - 10%, 3 PHASE, 50/60 HZ+ OR -2HZ, 8% LINE REGULATION COMMERCIAL POWER OR 100 KW GENERATORS USED BY AMEDD TOE AND OTHER KW; INSTANTANEOUS POWER DEMAND NOT EXCEEDING 75KW; SHAL NOT REQUIRE BETTER THAN 10% LINE VOLTAGE REGULATION WHEN OPERATING AT MAX LINE CURRENT SETTING;PRIMARY DISCONNECT MEANS FROM LOCATION READILY ACCESSIBLE BY OPERATOR; AMPACITY OF CIRCUIT BREAKER RATED AT LEAST 50% OF MONETARY MAX RATING OF EQUIP; SIZED TO ACCOMMODATE 2 AWG CONDUCTORS; COTS SCANNER AND ALL MAJOR COMPONENTS AND CERTIFIED 21 CFR REQUIREMENTS; ADDITIONAL COMPONETS TO FIT INSIDE ISO SHELTER FOR STORAGE AND USE MAY BE REQUIRED; C/O CT GANTRY; OPERATIOR CONTROL, PATIENT SUPPORT. SYSTEM CAPABLE OF BEING INSTALLED AND OPERATED IN ISO SHELTERS, GANTRY TILT OF +/- 25 DEG; PATIENT TABLE; DISPLAY CONSOLE; CT IMAGE OUTPUT FROM CT COMPUTER VIA OPTICAL DISK |
| PART NAME ASSIGNED BY CONTROLLING AGENCY | COMPUTED TOMOGRAPHY SCANNER FIELD 208 VAC 50-60HZ EA |