Part Details | PHOTOGRAPHIC FILM PROCESSING MACHINE
6525-01-512-1960 A machine designed for processing photographic film in continuous and automatic operation. It must have facilities for developing and fixing. It may or may not have facilities for washing and drying.
Alternate Parts: AT 2000, A/T 2000, LLH368308, LL-H36-8308, 6525-01-512-1960, 01-512-1960, 6525015121960, 015121960
| Supply Group (FSG) | NSN Assigned | NIIN | Item Name Code (INC) |
|---|---|---|---|
| 65 | AUG 26, 2003 | 01-512-1960 | 06110 ( PROCESSING MACHINE, PHOTOGRAPHIC FILM ) |
REFERENCE DRAWINGS & PICTURES
Cross Reference | NSN 6525-01-512-1960
| Part Number | Cage Code | Manufacturer |
|---|---|---|
| A/T 2000 | 53542 | AIR TECHNIQUES, INC. |
| LL-H36-8308 | 66736 | NAVAL MEDICAL MATERIAL SUPPORTCOMMAND DETACHMENT |
Technical Data | NSN 6525-01-512-1960
| Characteristic | Specifications |
|---|---|
| CURRENT TYPE | AC |
| VOLTAGE IN VOLTS | 115.0 |
| PHASE | SINGLE |
| FREQUENCY RATING | 60.0 HERTZ NOMINAL |
| SPECIFIC USE | FILM PROCESSING MACHINE |
| USAGE DESIGN | FILM |
| LOADING TYPE | N ANY ACCEPTABLE |
| LOADING CHARACTERISTIC | DAYLIGHT |
| INSTALLATION DESIGN | PORTABLE |
| TEMP CONTROLLED FEATURE | INCLUDED |
| SPECIAL FEATURES | 8 AMPS; THREE PRONG PLUG REQUIRED; 2.5 MINUTE ENDO QUICK LOOK FILM PROCESSING; 5.5 MINUTE ARCHIVAL FILM PROCESSING; WATER FLOE .5 GPM; REQUIRES 80 PSI MAX WATER PRESSURE WITH 30PSI MIN; WATER SOURCE MUST HAVE MANUAL SHUT OFF NEAR PROCESSOR; ADAPTERS TO CONNECT .5 INCH MPT OR .75 INCH MALE GARDEN HOSE FITTING; SUPPLIED TUBING AT REAR OF PROCESSOR IS 6 FEET LONG; VACUUM BREAKER NOT REQUIRED; 1 INCH AIR GAP BETWEEN WATER SUPPLY INLET AND TOP OF TANK; WATER TEMP 55 DEG F TO 80 DEG F; MIXING VALVE NOT REQUIRED IF WATER TEMP IS IN RANGE; 1.5 INCH PVC STANDPIPE CONNECTED TO A 1.5 INCH PVC TRAP; DO NOT DRAIN INTO COPPER OR BRASS |
| PART NAME ASSIGNED BY CONTROLLING AGENCY | MODEL A/T 2000 X-RAY FILM PROCESSOR |