NSN 6525-01-601-2327

Part Details | PHANTOM,TOMOGRAPHIC

6525-01-601-2327

Alternate Parts: 043762, 043-762, 6525-01-601-2327, 01-601-2327, 6525016012327, 016012327

Supply Group (FSG) NSN Assigned NIIN Item Name Code (INC)
REFERENCE DRAWINGS & PICTURES
Cross Reference | NSN 6525-01-601-2327
Part Number Cage Code Manufacturer
043-76213308BIODEX MEDICAL SYSTEMS, INC.DIV BIODEX
Technical Data | NSN 6525-01-601-2327
Characteristic Specifications
SPECIAL FEATURESPROVIDES CONSISTENT PERFORMANCE INFORMATION FOR ANY SINGLE PHOTON EMISSION COMPUTED TOMOGRAPHY (SPECT) OR POSITRON EMISSION TOMOGRAPHY (PET) SYSTEM; MULTIPLE PERFORMANCE CHARACTERISTICS OF CAMERA-BASED SPECT SYSTEMS ARE EVALUATED FROM A SINGLE SCAN OF THE PHANTOM; CYLINDER INTERIOR DIMENSIONS: 8.5 IN DIA X 7.32 IN H (21.6 X 18.6 CM); CYLINDER WALL THICKNESS: 0,125 IN (3.2 MM); COLD ROD DIAMETERS: 6.4, 7.9, 9.5, 11.1, 12.7, 16.0 MM; SOLID SPHERE DIAMETERS: 12.7, 15.9, 19.1, 25.4, 31.8, 38 MM
PART NAME ASSIGNED BY CONTROLLING AGENCYJASZCZAK SPECT PHANTOM, STANDARD