Part Details | CLINICAL CHEMISTRY ANALYZER
6630-01-332-7886 A microprocessor based instrument incorporating more than one method of analysis such as colorimetric and potentiometric/ION selective or kinetic enzyme. It may be either a stand-alone unit or modular in design. For instruments which have only one method of analysis see ANALYZER, as modified. Excludes all instruments which employ a chromatographic method of analysis.
Alternate Parts: YYJ8422172, J8422172, DT60WDTEDTSC, DT60-W/DTE/DTSC, DT60 II WDTE & DTSC, DT60 II W/DTE & DTSC, DT60 II, 842 2172, VITROS DT60 II, 6630-01-332-7886, 01-332-7886, 6630013327886, 013327886
| Supply Group (FSG) | NSN Assigned | NIIN | Item Name Code (INC) |
|---|---|---|---|
| 66 | JAN 17, 1991 | 01-332-7886 | 42054 ( ANALYZER, CLINICAL CHEMISTRY ) |
REFERENCE DRAWINGS & PICTURES
Cross Reference | NSN 6630-01-332-7886
| Part Number | Cage Code | Manufacturer |
|---|---|---|
| YYJ8422172 | KA068 | ALERE LIMITED |
| J8422172 | U8163 | AXIS-SHIELD DIAGNOSTICS LTD |
| DT60-W/DTE/DTSC | 19139 | EASTMAN KODAK COMPANYDBA KODAK |
| DT60 II W/DTE & DTSC | 03KD1 | ORTHO-CLINICAL DIAGNOSTICS, INC. |
| DT60 II | 03KD1 | ORTHO-CLINICAL DIAGNOSTICS, INC. |
| 842 2172 | 03KD1 | ORTHO-CLINICAL DIAGNOSTICS, INC. |
| VITROS DT60 II | 03KD1 | ORTHO-CLINICAL DIAGNOSTICS, INC. |
Technical Data | NSN 6630-01-332-7886
| Characteristic | Specifications |
|---|---|
| METHOD OF ANALYSIS | COLORIMETRIC AND KINETIC ENZYME/RATE TEST AND POTENTIOMETRIC/ION SELECTIVE ELECTRODE |
| WAVELENGTH RANGE DATA NON-AVAILABILITY | NOT INCLUDED AS AN ESSENTIAL CHARACTERISTIC IN THE SPECIFICATION/CID COVERING THIS STANDARD NSN |
| ELECTRODE TYPE | CALCIUM/CA++ AND CARBON DIOXIDE/CO2 AND CHLORIDE/CL- AND POTASSIUM/K+ AND SODIUM/NA+ |
| TEST TYPE FOR WHICH DESIGNED | ALBUMIN; AMYLASE; CREATININE; GLUCOSE; UREA NITROGEN; PHOSPHORUS; TOTAL BILIRUB; TOTAL PROTEIN; ASPARTATE AMINOTRANSFERASE,(AST); ALANINE AMINOTRANSFERASE,(ALT); CREATINE KINASE (CK); CALCIUM; SODIUM; POTASSIUM; CHLORIDE; CARBON DIOXIDE |
| ANALYSIS TEST/MEASUREMENT RANGE | ALBUMIN 3-5 G/DL; AMYLASE 5-900 U/L; CREATINE 0.1-15 MG/DL; GLUCOSE 20-450 MG/DL; UREA NITROGEN 1-100 MG/DL; PHOSPHORUS 0.5-13 MG/DL; TOTAL BILIRUBIN 0.1-20 MG/DL; TOTAL PROTEIN 2-11 G/DL; ASPARTATE AMINOTRANSFERASE (AST) 4-950 U/L; ALANINE AMINOTRANSFERASE (ALT) 3-950 U/L; CREATINEKINASE (CK) 20-1600 U/L; CALCIUM 3-15 MG/DL; SODIUM 95-215 MMO1/L; POTASSIUM 1-11 MMQ1/L; CHLORIDE 65-140 MMO1/L; CARBON DIOXIDE 5-50 MMO1/L |
| TEST SAMPLE/SPECIMEN SIZE | 12.0 MICROLITERS |
| THROUGHPUT PER HOUR | 50 SAMPLES/SPECIMENS MINIMUM |
| TEST SAMPLE/SPECIMEN DELIVERY DEVICE | PIPET, HANDHELD AUTOMATIC |
| FEATURES PROVIDED | BATCH PROCESSING AND DATA ENTRY KEYBOARD AND DISCRETE PROCESSING AND INTEGRAL PRINTER AND RS232 PORT AND STAT PROCESSING |
| CURRENT TYPE | AC |
| VOLTAGE IN VOLTS | B110.0$$B240.0 |
| FREQUENCY RATING | A50.0$A60.0 |
| HEIGHT | 7.000 INCHES |
| WIDTH | 42.000 INCHES |
| DEPTH | 14.000 INCHES |
| ACCESSORY COMPONENTS AND QUANTITY | PIPETTE(S) |
| SPECIAL FEATURES | DRY CHEMICAL TECHNOLOGY; SOLID STATE |