Part Details | THERMOMETER CALIBRATOR
6685-01-224-4282 An item specifically designed to determine by comparison the accuracy of calibration of thermometers.
Alternate Parts: MLEE82117B, MLEE82-117B, AA59419, A-A-59419, 7013, 5309, 6685-01-224-4282, 01-224-4282, 6685012244282, 012244282
| Supply Group (FSG) | NSN Assigned | NIIN | Item Name Code (INC) |
|---|---|---|---|
| 66 | JAN 16, 1986 | 01-224-4282 | 60161 ( CALIBRATOR, THERMOMETER ) |
REFERENCE DRAWINGS & PICTURES
Cross Reference | NSN 6685-01-224-4282
| Part Number | Cage Code | Manufacturer |
|---|---|---|
| MLEE82-117B | 30653 | AIR FORCE METROLOGY AND CALIBRATION |
| A-A-59419 | 58536 | FEDERAL COMMERCIAL ITEM DESCRIPTIONPROMULGATED BY GENERAL SERVICES |
| 7013 | 64841 | HART SCIENTIFIC LLC |
| 5309 | 64841 | HART SCIENTIFIC LLC |
Technical Data | NSN 6685-01-224-4282
| Characteristic | Specifications |
|---|---|
| CALIBRATION TYPE | R ANY ACCEPTABLE AND Q ANY ACCEPTABLE |
| INSTRUMENT CALIBRATION RANGE | COLD BATH -100 TO +70 DEG C (-148 TO +158 DEG F); HOT BATH +60 TO +300 DEG C (+140 TO +572 DEG F) |
| CALIBRATION ACCURACY RATING | -100 TO -40 DEG C - PORM 5 DEG C; -40 TO +150 DEG C - PORM 1 DEG C; +150 TO +300 DEG C - PORM 5 DEG C |
| AC VOLTAGE RATING | 208.0 VOLTS MINIMUM AND 230.0 VOLTS MAXIMUM |
| FREQUENCY RATING | 60.0 HERTZ NOMINAL |
| PHASE | SINGLE |
| OVERALL WIDTH | 45.250 INCHES NOMINAL |
| OVERALL HEIGHT | 45.000 INCHES NOMINAL |
| OVERALL DEPTH | 26.250 INCHES NOMINAL |
| MOUNTING CONFIGURATION | FREE STANDING CABINET ON CASTORS |
| SPECIAL FEATURES | C/O TWO TEST WELLS, EACH WELL IS 2.6"DIA X 11/3" DEEP (60 CUBIC INCHES) AND HAVE A 0.25" DEEP MERCURY TRAP. TRANSFER FLUIDS: COLD BATH - HALOCARBON 0.8; HOT BATH - DOW 710. |
| UNPACKAGED UNIT WEIGHT | 450.000 POUNDS |
| FUNCTIONAL DESCRIPTION | CONSISTS OF 2 INDEPENDENTLY OPERABLE BATH SYSTEMS. EACH BATH HAS ITS OWN TANK WITH A CONSTANT LEVEL TEST WELL & PUMP. EACH HAS ITS OWN TEMPERATURE CONTROLLER, PROBE, & HEATING AND COLLING SYSTEMS. SEPERATE FRONT PANEL POWER CONTROLS, LABELED HOT BATH AND COLD BATH, ALLOW INDEPENDENT POWER UP AND USE. |
| REFERENCE DATA AND LITERATURE | COMMERCIAL MANUAL |